Sun

Features:

  • Horizontal axis
  • Under ISO 5 laminar flow, ISO 8 cleanroom
  • Effective diameter: 550 mm
  • Effective length: 605 mm
  • Interface plates:
  1.  base plate: 250 x 540 mm
  2. removable top plate: 496 x 540 mm

Acquisitions, equipment:

  • Pressure: 1 Piezo gauge, 1 Pirani gauge, 2 Penning gauges
  • Temperature: 8 sensors, type: 100 Ω platinum (4-wire measurement)
  • Residual Gas Analyzer (RGA) with Differential Pumping
  • Clock in UTC (Coordinated Universal Time)
  • Write to ASCII files every second

Pumping:

  •   1 Primary Pump
  •   1 Turbomolecular pump
  •   Vacuum pressure: 1 × 10⁻⁷ hPa

Drying:

  • Radiative drying
  • Components mounted on a stainless steel plate resting on 4 feet isolated from the screens
  • Maximum temperature: 200 °C, maximum rate of change of 1 °C per minute on the screens
  • Three screens controlled simultaneously:

 cylinder = "center" (master programmable controller)

 2 discs: “front” and “rear” (slave regulators)

  • 2 safety features for the heating power shut-off:

Pressure > 2 × 10⁻⁴ hPa (10-second delay)

Measurement of one of the three controllers > alarm threshold (programmable in consultation with the experimenters; typically: 5°C above the drying oven setpoint)

  • No cooling system for returning to room temperature:

Allow 36 hours for drying at 100°C and 48 hours for drying at 200°C

  • No electrical feedthroughs are available for experimenters
Scroll up