Sun
Features:
- Horizontal axis
- Under ISO 5 laminar flow, ISO 8 cleanroom
- Effective diameter: 550 mm
- Effective length: 605 mm
- Interface plates:
- base plate: 250 x 540 mm
- removable top plate: 496 x 540 mm
Acquisitions, equipment:
- Pressure: 1 Piezo gauge, 1 Pirani gauge, 2 Penning gauges
- Temperature: 8 sensors, type: 100 Ω platinum (4-wire measurement)
- Residual Gas Analyzer (RGA) with Differential Pumping
- Clock in UTC (Coordinated Universal Time)
- Write to ASCII files every second
Pumping:
- 1 Primary Pump
- 1 Turbomolecular pump
- Vacuum pressure: 1 × 10⁻⁷ hPa
Drying:
- Radiative drying
- Components mounted on a stainless steel plate resting on 4 feet isolated from the screens
- Maximum temperature: 200 °C, maximum rate of change of 1 °C per minute on the screens
- Three screens controlled simultaneously:
cylinder = "center" (master programmable controller)
2 discs: “front” and “rear” (slave regulators)
- 2 safety features for the heating power shut-off:
Pressure > 2 × 10⁻⁴ hPa (10-second delay)
Measurement of one of the three controllers > alarm threshold (programmable in consultation with the experimenters; typically: 5°C above the drying oven setpoint)
- No cooling system for returning to room temperature:
Allow 36 hours for drying at 100°C and 48 hours for drying at 200°C
- No electrical feedthroughs are available for experimenters